MODEL HR-100A / HR-100AZ HR-150A / HR-150AZ HR-250A / HR-250AZ
Weighing capacity 102g 152g 252g
Minimum weighing value 0.1mg

Repeatability/Std Dev

0.01mg 0 to 200 g: 0.1 mg
200 to 252 g: 0.2 mg
Linearity ±0.2mg ±0.3mg
Stabilization Time (set to FAST) Approx. 2 second*
Sensitivity Drift ±2ppm/°C (10°C-30°C/50°F-86°F)
Calibration HR-A External
HR-AZ Internal
Clock and calendar function HR-A No
HR-AZ Yes
Operating temperature 5 to 40°C (41 to 104°F), 85%RH or less (non-condensing)
Display Refresh rate 5 times / second or 10 times / second, selectable
Measurement Units (Display mode) g, mg, oz, ozt, ct, momme, dwt, GN, tael, tola, 
pieces (counting mode), % (percent mode), DS (density mode), user definable functions
Counting mode Minimum unit mass = 0.1mg / Number of samples = 10, 25, 50 or 100 pieces
Percent mode Minimum 100% ref. mass = 10.0 mg / Minimum 100% display = 0.01%, 0.1%, or 1% (depends on the reference mass stored)
Standard interface RS-232C standard (see Accessories for more interface options)
Applicable calibration weight value 100g, 50g 150g, 100g, 50g 250g, 200g, 100g, 50g
Weighing pan diameter Ø90 mm
External dimensions 198 (W) x 294 (D) x 315 (H) mm / 8 (W) x 12 (D) x 12.4 (H) in
Net weight HR-A 3.5 kg / 7.8 lb
HR-AZ 3.9 kg / 8.9 lb
Power supply AC adapter
Power consumption Approx. 11 VA (supplied by the AC adapter)

 